2h Biological Indicator for Hydrogen Peroxide (FDA)
More info about this product
Additional information
Brand | Bionova |
---|---|
Process | Hydrogen Peroxide |
Packaging | 50 |
Microorganism | Geobacillus stearothermophilus (ATCC 7953) |
Population | 10^6 Spores/carrier |
Read-Out Time | 2 h |
Regulations | IRAM 37102-1, ISO 11138-1 |
Description
Rapid Self-contained Biological Indicator for Plasma or Vaporized Hydrogen Peroxide sterilization processes.Geobacillus stearothermophilus (ATCC 7953) 10^6 spores per vial. Readout: 2 hours.
Terragene® Bionova® BT95 Fluorescence Rapid Readout Biological Indicators are single-use Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier and a glass ampoule with a culture medium, enclosed with a colored cap.
Each tube contains a population of Geobacillus stearotermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT95 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide.
Instructions for Use
1. Identify the Terragene® Bionova® BT95 SCBI by writing the sterilizer number (in case of having more than one), load number and processing date on the label.
2. Pack the SCBI along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas which are considered most inaccessible for the sterilizing agent (e.g., the center of the load and areas near the door).
3. Sterilize as usual.
4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the SCBI from the package. CAUTION: Wear safety glasses and gloves when removing Terragene® Bionova® BT95 SCBI from the sterilized package.
WARNING: Do not crush or handle the SCBI excessively, since this might cause the glass ampoule to burst. Let the SCBI cool down until it reaches room temperature.
5. Check the Process Indicator on SCBI label. A color change to green indicates that the SCBI has been exposed to Hydrogen Peroxide.
IMPORTANT: This color change does not evidence the process effectiveness to achieve sterility. If the Process Indicator color has not changed, check the sterilization process.
6. Press the lid to seal the tube. Crush the ampoule contained in the SCBI with an individual ampoule crusher or with the ampoule crusher placed within the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature in a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the SCBI in the incubator.
IMPORTANT: Use a non-sterilized SCBI as a positive control in order to ensure that correct incubation conditions were met; the capability of medium to promote rapid growth; that viability of spores has not been altered due to improper storage temperature, humidity or proximity to chemicals and proper functioning of Terragene® Bionova® Auto-Reader Incubators. Both, the positive control indicator and the processed indicator, should belong to the same batch.
7. Incubate the processed indicator and the positive control indicator in the appropriate Terragene® Bionova® Auto-Reader incubators for a maximum of 2 hours at (60 ± 2) °C for rapid readout.
NOTE: Holding time between sterilization and incubation should not exceed a 7-day period. Fluorescence detection by the reader (excitation 340-380 nm / emission 455-465 nm) means failure in the sterilization process. If no fluorescence is detected at 2-hour incubation, the result is negative.
Description
Rapid Self-contained Biological Indicator for Plasma or Vaporized Hydrogen Peroxide sterilization processes.Geobacillus stearothermophilus (ATCC 7953) 10^6 spores per vial. Readout: 2 hours.
Terragene® Bionova® BT95 Fluorescence Rapid Readout Biological Indicators are single-use Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier and a glass ampoule with a culture medium, enclosed with a colored cap.
Each tube contains a population of Geobacillus stearotermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT95 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide.
Instructions for Use
1. Identify the Terragene® Bionova® BT95 SCBI by writing the sterilizer number (in case of having more than one), load number and processing date on the label.
2. Pack the SCBI along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas which are considered most inaccessible for the sterilizing agent (e.g., the center of the load and areas near the door).
3. Sterilize as usual.
4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the SCBI from the package. CAUTION: Wear safety glasses and gloves when removing Terragene® Bionova® BT95 SCBI from the sterilized package.
WARNING: Do not crush or handle the SCBI excessively, since this might cause the glass ampoule to burst. Let the SCBI cool down until it reaches room temperature.
5. Check the Process Indicator on SCBI label. A color change to green indicates that the SCBI has been exposed to Hydrogen Peroxide.
IMPORTANT: This color change does not evidence the process effectiveness to achieve sterility. If the Process Indicator color has not changed, check the sterilization process.
6. Press the lid to seal the tube. Crush the ampoule contained in the SCBI with an individual ampoule crusher or with the ampoule crusher placed within the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature in a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the SCBI in the incubator.
IMPORTANT: Use a non-sterilized SCBI as a positive control in order to ensure that correct incubation conditions were met; the capability of medium to promote rapid growth; that viability of spores has not been altered due to improper storage temperature, humidity or proximity to chemicals and proper functioning of Terragene® Bionova® Auto-Reader Incubators. Both, the positive control indicator and the processed indicator, should belong to the same batch.
7. Incubate the processed indicator and the positive control indicator in the appropriate Terragene® Bionova® Auto-Reader incubators for a maximum of 2 hours at (60 ± 2) °C for rapid readout.
NOTE: Holding time between sterilization and incubation should not exceed a 7-day period. Fluorescence detection by the reader (excitation 340-380 nm / emission 455-465 nm) means failure in the sterilization process. If no fluorescence is detected at 2-hour incubation, the result is negative.

Related Products
Latest News
FAQs
You can get commercial assistance contacting us here
You can find product resources and information, visiting our download center here or on our products page.
Our products can be obtained by talking with our commercial executives, that will guide through all your buying processes, giving our best effort for an easy and comforting purchasing experience. You can contact our team here.
To be a Terragene distributor, you can get in touch with our commercial executives filling our contact formulary here.
You can download the updated softwares here.