Customize Consent Preferences

We use cookies to help you navigate efficiently and perform certain functions. You will find detailed information about all cookies under each consent category below.

The cookies that are categorized as "Necessary" are stored on your browser as they are essential for enabling the basic functionalities of the site. ... 

Always Active

Necessary cookies are required to enable the basic features of this site, such as providing secure log-in or adjusting your consent preferences. These cookies do not store any personally identifiable data.

No cookies to display.

Functional cookies help perform certain functionalities like sharing the content of the website on social media platforms, collecting feedback, and other third-party features.

No cookies to display.

Analytical cookies are used to understand how visitors interact with the website. These cookies help provide information on metrics such as the number of visitors, bounce rate, traffic source, etc.

No cookies to display.

Performance cookies are used to understand and analyze the key performance indexes of the website which helps in delivering a better user experience for the visitors.

No cookies to display.

Advertisement cookies are used to provide visitors with customized advertisements based on the pages you visited previously and to analyze the effectiveness of the ad campaigns.

No cookies to display.

Skip to content

2h Biological Indicator for LTSF

Additional information

Brand

Bionova

Process

Formaldehyde

Packaging

50

Microorganism

Geobacillus stearothermophilus (ATCC® 7953)

Population

10^6 Spores/carrier

Read-Out Time

2 h

Regulations

IRAM 37102-1, ISO 11138-1, ISO11138-5

Field of activity

Healthcare

Description

BT102 / Rapid Self-contained Biological Indicator for Formaldehyde sterilization processes. Geobacillus stearothermophilus (ATCC® 7953) 10^6 spores per vial. Readout: 2 hours.

Characteristics

Polypropylene tube: 50.7 mm high x 8.5 mm external diameter. Wall thickness: 0.5 mm.

Polypropylene cap: 16.4 mm high x 10.7 mm external diameter. Wall thickness: 0.9 mm.

Cap filter: polyethylene fibers, 17.0 mm diameter.

Glass ampoule: 35.0 – 40.0 mm high. mm. External diameter: 6.8 mm.

Wall thickness: 0.2 – 0.3 mm.

Culture medium 0.5 – 0.7 ml, purple color.

Polypropylene microfiber on spore carrier, 17.0 mm diameter.

Spores carrier: polyethylene fibers, 17.0 mm of diameter. 10^6 Geobacillus stearothermophilus ATCC® 7953 spores per vial.

Instructions for use

1. Identify the Bionova® BT102 Biological Indicator by writing the sterilizer number (in case of having more than one), load number and processing date on the label.

2. Pack the biological indicator along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas considered a priori most inaccessible for the sterilizing agent (Formaldehyde). Generally, the center of the load and areas near the door are more problematic.

3. Sterilize as usual.

4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the biological indicator from the package. CAUTION: Wear safety glasses and gloves when removing the Bionova® BT102 Biological Indicator from the sterilized package.

WARNING: Do not crush or handle the biological indicator excessively, since this might cause the glass ampoule to burst.

5. Let the biological indicator cool down until it reaches room temperature.

6. Check the chemical indicator on the label of the biological indicator. A color change to green indicates that the biological indicator has been exposed to Formaldehyde.

IMPORTANT: this color change does not evidence the process effectiveness to achieve sterility. If the chemical indicator color has not changed, check the sterilization process.

7. Press the cap to seal the tube. Crush the ampoule contained in the biological indicator with an individual ampoule crusher or with the ampoule crusher placed within the top of the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature of a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the biological indicator in the incubator.

Description

BT102 / Rapid Self-contained Biological Indicator for Formaldehyde sterilization processes. Geobacillus stearothermophilus (ATCC® 7953) 10^6 spores per vial. Readout: 2 hours.

Characteristics

Polypropylene tube: 50.7 mm high x 8.5 mm external diameter. Wall thickness: 0.5 mm.

Polypropylene cap: 16.4 mm high x 10.7 mm external diameter. Wall thickness: 0.9 mm.

Cap filter: polyethylene fibers, 17.0 mm diameter.

Glass ampoule: 35.0 – 40.0 mm high. mm. External diameter: 6.8 mm.

Wall thickness: 0.2 – 0.3 mm.

Culture medium 0.5 – 0.7 ml, purple color.

Polypropylene microfiber on spore carrier, 17.0 mm diameter.

Spores carrier: polyethylene fibers, 17.0 mm of diameter. 10^6 Geobacillus stearothermophilus ATCC® 7953 spores per vial.

Instructions for use

1. Identify the Bionova® BT102 Biological Indicator by writing the sterilizer number (in case of having more than one), load number and processing date on the label.

2. Pack the biological indicator along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas considered a priori most inaccessible for the sterilizing agent (Formaldehyde). Generally, the center of the load and areas near the door are more problematic.

3. Sterilize as usual.

4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the biological indicator from the package. CAUTION: Wear safety glasses and gloves when removing the Bionova® BT102 Biological Indicator from the sterilized package.

WARNING: Do not crush or handle the biological indicator excessively, since this might cause the glass ampoule to burst.

5. Let the biological indicator cool down until it reaches room temperature.

6. Check the chemical indicator on the label of the biological indicator. A color change to green indicates that the biological indicator has been exposed to Formaldehyde.

IMPORTANT: this color change does not evidence the process effectiveness to achieve sterility. If the chemical indicator color has not changed, check the sterilization process.

7. Press the cap to seal the tube. Crush the ampoule contained in the biological indicator with an individual ampoule crusher or with the ampoule crusher placed within the top of the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature of a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the biological indicator in the incubator.