Indicator for ultrasonic cavitation performance tests – CDWU-Z
Additional information
Brand | Chemdye |
---|---|
Process | Cleaning and Thermodisinfection |
Packaging | 30, 6 |
Regulations | ISO 13485:2016 / NS-EN ISO 13485:2016. |
Description
CDWU-Z / Designed for ultrasonic cleaning processes routine control and monitoring ultrasonic cavitation effectiveness of ultrasonic washing machines operating at frequencies equal to or greater than 35 kHz, at temperatures ranging from 18 °C to 70 °C.
Characteristics
-2 ml clear glass vial: 31.0 mm high x 11.0 mm external diameter, 8.0 mm neck diameter. Wall thickness: 0.5 mm.
-Brown/Red rubber septum: 8.0 x 1.0 mm.
-Aluminum top cap: 5.5 mm high x 11.7 mm diameter.
-Wall thickness: 0.3 mm.
-Ceramic cylinders: 2.5 mm high and 2.5 mm diameter.
-Blue clear reactive solution: 1.5 ml.
-Allows detection of deficiencies in the cavitation performance of the ultrasonic washing machine. A self-adhesive Record Card accompanies each vial where data about the ultrasonic washing control process can be recorded.
-Challenge level: high.
-Initial Color: blue.
-Result: yellow transparent solution.
Instructions for use
Ultrasonic washers should be checked at least once per week and the results should be recorded. There are two types of tests (see figure) and CDWU-Z indicator can be used in both:
-Periodic functional test: for initial set-up of the washing machine, and at regular intervals of three months or after a repair. The purpose of this test is to verify the uniform distribution of the cavitation process in the washer tank.
-Test routine: is performed weekly or daily, to verify that the cavitation works correctly. This test allows detecting a possible failure in the performance of the ultrasonic washing machine.
Description
CDWU-Z / Designed for ultrasonic cleaning processes routine control and monitoring ultrasonic cavitation effectiveness of ultrasonic washing machines operating at frequencies equal to or greater than 35 kHz, at temperatures ranging from 18 °C to 70 °C.
Characteristics
-2 ml clear glass vial: 31.0 mm high x 11.0 mm external diameter, 8.0 mm neck diameter. Wall thickness: 0.5 mm.
-Brown/Red rubber septum: 8.0 x 1.0 mm.
-Aluminum top cap: 5.5 mm high x 11.7 mm diameter.
-Wall thickness: 0.3 mm.
-Ceramic cylinders: 2.5 mm high and 2.5 mm diameter.
-Blue clear reactive solution: 1.5 ml.
-Allows detection of deficiencies in the cavitation performance of the ultrasonic washing machine. A self-adhesive Record Card accompanies each vial where data about the ultrasonic washing control process can be recorded.
-Challenge level: high.
-Initial Color: blue.
-Result: yellow transparent solution.
Instructions for use
Ultrasonic washers should be checked at least once per week and the results should be recorded. There are two types of tests (see figure) and CDWU-Z indicator can be used in both:
-Periodic functional test: for initial set-up of the washing machine, and at regular intervals of three months or after a repair. The purpose of this test is to verify the uniform distribution of the cavitation process in the washer tank.
-Test routine: is performed weekly or daily, to verify that the cavitation works correctly. This test allows detecting a possible failure in the performance of the ultrasonic washing machine.

Related Products
More Products
Latest News
FAQs
You can get commercial assistance contacting us here
You can find product resources and information, visiting our download center here or on our products page.
Our products can be obtained by talking with our commercial executives, that will guide through all your buying processes, giving our best effort for an easy and comforting purchasing experience. You can contact our team here.
To be a Terragene distributor, you can get in touch with our commercial executives filling our contact formulary here.
You can download the updated softwares here.