Customize Consent Preferences

We use cookies to help you navigate efficiently and perform certain functions. You will find detailed information about all cookies under each consent category below.

The cookies that are categorized as "Necessary" are stored on your browser as they are essential for enabling the basic functionalities of the site. ... 

Always Active

Necessary cookies are required to enable the basic features of this site, such as providing secure log-in or adjusting your consent preferences. These cookies do not store any personally identifiable data.

No cookies to display.

Functional cookies help perform certain functionalities like sharing the content of the website on social media platforms, collecting feedback, and other third-party features.

No cookies to display.

Analytical cookies are used to understand how visitors interact with the website. These cookies help provide information on metrics such as the number of visitors, bounce rate, traffic source, etc.

No cookies to display.

Performance cookies are used to understand and analyze the key performance indexes of the website which helps in delivering a better user experience for the visitors.

No cookies to display.

Advertisement cookies are used to provide visitors with customized advertisements based on the pages you visited previously and to analyze the effectiveness of the ad campaigns.

No cookies to display.

Skip to content

Chemical indicator for Helix-PCD systems

Additional information

Brand

Chemdye

Process

Ethylene Oxide

Packaging

250

Initial Color

Light Blue

Final Color

Green

Indicator Type

Type 2 Chemical Indicator

Regulations

ISO 11140-1

Field of activity

Healthcare

Description

PCD-A-E1 / Test strips for use in Helix-PCD System. Laminated self-adhesive strips printed with indicator ink (77 x 6 mm), for Ethylene Oxide sterilization processes. 240 min. 54° C, 600 mg/l 60% RH. Turns from Light Blue to Green. Metals free. ISO 11140-1 Type 2.

Chemical strips (Type 2 Chemical Indicator according to ISO 11140-1:2014) for use with Helix-PCD in the control of ethylene oxide sterilization processes. Indicator 100% free of toxic heavy metals.

Instructions for use

1. Insert PCD-A-E1 chemical indicator, according to the selected process conditions, within the pin (metal capsule) or slot (plastic capsule) of the screw cap; settle the reactive indicating ink towards the capsule lumen, avoiding folding those areas printed with reactive indicating ink.

IMPORTANT: avoid contact between PCD-A-E1 chemical indicator and capsule interior surface, because it can alter the result.

2. Put the capsule cap with moderate pressure.

3. Settle the system into the fabric bag.

4. The entire system should be sterilized according to the usual cycle used. For example: 240 minutes, 600 mg/l, 54 ºC, 60% RH. Do not use cycles below 90 minutes and EO 300 mg/l.

Initial color: sky blue

Final color: green

5. After the period of ventilation, remove the indicator and evaluate the result. If the 4 bars change to the final color, the result can be considered successful, which means the sterilizer is working correctly. If one bar doesn€™t reach the final color, it means that residual air is present in the sterilizer, therefore a malfunction of the equipment.

Description

PCD-A-E1 / Test strips for use in Helix-PCD System. Laminated self-adhesive strips printed with indicator ink (77 x 6 mm), for Ethylene Oxide sterilization processes. 240 min. 54° C, 600 mg/l 60% RH. Turns from Light Blue to Green. Metals free. ISO 11140-1 Type 2.

Chemical strips (Type 2 Chemical Indicator according to ISO 11140-1:2014) for use with Helix-PCD in the control of ethylene oxide sterilization processes. Indicator 100% free of toxic heavy metals.

Instructions for use

1. Insert PCD-A-E1 chemical indicator, according to the selected process conditions, within the pin (metal capsule) or slot (plastic capsule) of the screw cap; settle the reactive indicating ink towards the capsule lumen, avoiding folding those areas printed with reactive indicating ink.

IMPORTANT: avoid contact between PCD-A-E1 chemical indicator and capsule interior surface, because it can alter the result.

2. Put the capsule cap with moderate pressure.

3. Settle the system into the fabric bag.

4. The entire system should be sterilized according to the usual cycle used. For example: 240 minutes, 600 mg/l, 54 ºC, 60% RH. Do not use cycles below 90 minutes and EO 300 mg/l.

Initial color: sky blue

Final color: green

5. After the period of ventilation, remove the indicator and evaluate the result. If the 4 bars change to the final color, the result can be considered successful, which means the sterilizer is working correctly. If one bar doesn€™t reach the final color, it means that residual air is present in the sterilizer, therefore a malfunction of the equipment.