Thermodisinfection Indicator for moist heat disinfection

Additional information
Brand

Integron

Process

Cleaning and Thermodisinfection

Packaging

200

Initial Color

Green

Final Color

Purple

Regulations

ISO 13485, ISO 15883-1

Field of activity

Dental, Healthcare, Pharmaceutical

Description

IT27W-5 / One-side-laminated strip (70 x 30 mm) printed with indicator ink for routine control of thermal disinfection processes, 5 min. 90° C, A0=3000. Turns from Green to Purple.

After cleaning, surgical instruments must be disinfected to ensure safe subsequent handling and processing. Thermal disinfection with moist heat, is the most common method for disinfection of medical devices in the hospital setting.

Disinfection effects can be defined by the A0 value, a physical parameter specified in ISO 15883-1, which denotes the inactivation of microorganisms. A particular A0 value can be achieved with the most diverse temperature/time combinations. For processing surgical instruments, an A0 value of 600 is specified in ISO 15883-2, corresponding to a hold time of 1 minute at 90 °C, ensuring lethality of bacteria, mycobacteria, fungi and heat-sensitive viruses.

If efficacy is also to be ensured against heat-resistant viruses, e.g. hepatitis B, many authorities in the field of disinfection, like the Robert Koch Institute, recommend an A0 value of 3000, corresponding to a temperature of 90 °C with a hold time of 5 min. A higher safety approach from this institute comprises a cycle at 93 ºC, 10 minutes, a program known as the €œBGA/RKI or epidemic program.

Integron® Indicators have been designed to react in moist heat disinfection processes in washer-disinfectors, providing an accurate, convenient method of routine control, and assuring disinfection efficiency. The green indicating ink was developed to turn to purple when temperature and time conditions are met.

Indications for use

Use the correspondent Integron® Indicator for monitoring thermal disinfection step in washer-disinfectors according to the following programs:

IT27W-1: 90 ºC, 1 minute. A0 = 600

IT27W-5: 90 ºC, 5 minutes. A0 =3000

IT27W-10: 93 ºC, 10 minutes. BGA/RKI/Epidemic Program

Description

IT27W-5 / One-side-laminated strip (70 x 30 mm) printed with indicator ink for routine control of thermal disinfection processes, 5 min. 90° C, A0=3000. Turns from Green to Purple.

After cleaning, surgical instruments must be disinfected to ensure safe subsequent handling and processing. Thermal disinfection with moist heat, is the most common method for disinfection of medical devices in the hospital setting.

Disinfection effects can be defined by the A0 value, a physical parameter specified in ISO 15883-1, which denotes the inactivation of microorganisms. A particular A0 value can be achieved with the most diverse temperature/time combinations. For processing surgical instruments, an A0 value of 600 is specified in ISO 15883-2, corresponding to a hold time of 1 minute at 90 °C, ensuring lethality of bacteria, mycobacteria, fungi and heat-sensitive viruses.

If efficacy is also to be ensured against heat-resistant viruses, e.g. hepatitis B, many authorities in the field of disinfection, like the Robert Koch Institute, recommend an A0 value of 3000, corresponding to a temperature of 90 °C with a hold time of 5 min. A higher safety approach from this institute comprises a cycle at 93 ºC, 10 minutes, a program known as the €œBGA/RKI or epidemic program.

Integron® Indicators have been designed to react in moist heat disinfection processes in washer-disinfectors, providing an accurate, convenient method of routine control, and assuring disinfection efficiency. The green indicating ink was developed to turn to purple when temperature and time conditions are met.

Indications for use

Use the correspondent Integron® Indicator for monitoring thermal disinfection step in washer-disinfectors according to the following programs:

IT27W-1: 90 ºC, 1 minute. A0 = 600

IT27W-5: 90 ºC, 5 minutes. A0 =3000

IT27W-10: 93 ºC, 10 minutes. BGA/RKI/Epidemic Program