Customize Consent Preferences

We use cookies to help you navigate efficiently and perform certain functions. You will find detailed information about all cookies under each consent category below.

The cookies that are categorized as "Necessary" are stored on your browser as they are essential for enabling the basic functionalities of the site. ... 

Always Active

Necessary cookies are required to enable the basic features of this site, such as providing secure log-in or adjusting your consent preferences. These cookies do not store any personally identifiable data.

No cookies to display.

Functional cookies help perform certain functionalities like sharing the content of the website on social media platforms, collecting feedback, and other third-party features.

No cookies to display.

Analytical cookies are used to understand how visitors interact with the website. These cookies help provide information on metrics such as the number of visitors, bounce rate, traffic source, etc.

No cookies to display.

Performance cookies are used to understand and analyze the key performance indexes of the website which helps in delivering a better user experience for the visitors.

No cookies to display.

Advertisement cookies are used to provide visitors with customized advertisements based on the pages you visited previously and to analyze the effectiveness of the ad campaigns.

No cookies to display.

Skip to content

Chemical indicator for Helix-PCD systems

Additional information

Brand

Chemdye

Process

Hydrogen Peroxide

Packaging

100

Initial Color

Purple

Final Color

Green

Indicator Type

Type 2 Chemical Indicator

Regulations

ISO 11140-1

Field of activity

Dental, Healthcare

Description

PCD-A-P1 / Test strips for use in Helix-PCD System. Self-adhesive varnished strips printed with indicator ink (77 x 6 mm), for Plasma or Vaporized Hydrogen Peroxide sterilization processes. Turns from Purple to Green. Metals free. ISO 11140-1 Type 2.

Chemical Strips (Type 2 Chemical Indicator according to ISO 11140-1) for

use in Helix-PCD for Hydrogen Peroxide.

Instructions for use

1-A. Remove the threaded cap from the metal or plastic capsule. Only for metal capsule Helix devices, remove the bolt attached to the cap (the bolt should NOT be used in Hydrogen Peroxide processes).

1-B. Place de chemical indicator in the capsule. Settle the reactive ink toward the capsule lumen, avoiding folds into the chemical ink.

IMPORTANT: Avoid the contact between the chemical indicator and the capsule internal surface, since the result can be altered.

2. Place the lid of the capsule with moderate pressure. ATTENTION! Do not place the system inside the cloth bag, since it can interfere with the process. The cloth bag must be used in steam processes.

NOTE: Terragene® suggests using the system Chemdye® Helix-PCD designed for small sterilizers (CDPCD2X025).

3.Start the selected program and test the process.

Initial color: purple

Final color: green

4. After the period of ventilation, remove the indicator and evaluate the result. If the 4 bars changed to the final color, the result is successful, the sterilizer is working correctly. If one bar is different to the final color, it shows the presence of residual air in the sterilizer, therefore an equipment malfunction.

We suggest retrying the test once again, before sending the equipment to technical service.

Description

PCD-A-P1 / Test strips for use in Helix-PCD System. Self-adhesive varnished strips printed with indicator ink (77 x 6 mm), for Plasma or Vaporized Hydrogen Peroxide sterilization processes. Turns from Purple to Green. Metals free. ISO 11140-1 Type 2.

Chemical Strips (Type 2 Chemical Indicator according to ISO 11140-1) for

use in Helix-PCD for Hydrogen Peroxide.

Instructions for use

1-A. Remove the threaded cap from the metal or plastic capsule. Only for metal capsule Helix devices, remove the bolt attached to the cap (the bolt should NOT be used in Hydrogen Peroxide processes).

1-B. Place de chemical indicator in the capsule. Settle the reactive ink toward the capsule lumen, avoiding folds into the chemical ink.

IMPORTANT: Avoid the contact between the chemical indicator and the capsule internal surface, since the result can be altered.

2. Place the lid of the capsule with moderate pressure. ATTENTION! Do not place the system inside the cloth bag, since it can interfere with the process. The cloth bag must be used in steam processes.

NOTE: Terragene® suggests using the system Chemdye® Helix-PCD designed for small sterilizers (CDPCD2X025).

3.Start the selected program and test the process.

Initial color: purple

Final color: green

4. After the period of ventilation, remove the indicator and evaluate the result. If the 4 bars changed to the final color, the result is successful, the sterilizer is working correctly. If one bar is different to the final color, it shows the presence of residual air in the sterilizer, therefore an equipment malfunction.

We suggest retrying the test once again, before sending the equipment to technical service.