Spore strips for Dry Heat, Steam and EtO (kit test)
More info about this product
Additional information
Brand | Bionova |
---|---|
Process | Dry Heat, Ethylene Oxide, Steam |
Packaging | 25 |
Microorganism | Bacillus atrophaeus (ATCC® 9372), Geobacillus stearothermophilus (ATCC® 7953) |
Population | 10^6 Spores/carrier |
Read-Out Time | 24 h, 48 h |
Regulations | IRAM 37102-1, IRAM 37102-2, IRAM 37102-3, ISO 11138-1, ISO 11138-2, ISO 11138-3, ISO 11138-4 |
Description
BT60-3 / Kit for Hospital test. Dual strips with Biological Indicator Spores for Steam, Ethylene Oxide and Dry Heat sterilization processes. Geobacillus stearothermophilus (ATCC® 7953) 10 ^ 5 or 10 ^ 6 and Bacillus atrophaeus (ATCC® 9372) 10 ^ 6 spores per strip.
Each Kit contains 3 Bionova BT60 Strip with Dual Spores.
Characteristics
The kit contains 1 Control Strip and 2 Test Strips.
25 mm x 70 mm non-absorbent material package, Violet code
5.0 x 25.0 mm filter paper strips.
10^5 or 10^6 Geobacillus stearothermophilus ATCC® 7953 spores and 10^6 Bacillus atrophaeus ATCC® 9372 spores per strip. Upon completion of the sterilization cycle, the spore strip should be tested in a suitable culture medium at 37 °C or at 60 °C.
We suggest the use of Bionova® MC1020-2 or MC1030-2 culture media to final reading in 48 hours at 37 °C for Ethylene Oxide and Dry Heat sterilization process control, or Bionova® MC20-2 or MC1020-2 culture media to final reading in 24 hours at 60 °C for Steam sterilization process control. For conventional culture media like TSB, incubation for 7 days is recommended.
D-ValueSTEAM: Not lower than 1.5 minutes. Conditions: 121 ºC.
Z-ValueSTEAM: Not lower than 6 ºC.
D-ValueEO: Not lower than 2.0 minutes. Conditions: 600 mg/l EO (100%), 60 % RH, 54 °C.
D-ValueDRY HEAT: Not lower than 2.0 minutes. Conditions: 160 °C.
Z-ValueDRY HEAT: Not lower than 20 °C.
Description
BT60-3 / Kit for Hospital test. Dual strips with Biological Indicator Spores for Steam, Ethylene Oxide and Dry Heat sterilization processes. Geobacillus stearothermophilus (ATCC® 7953) 10 ^ 5 or 10 ^ 6 and Bacillus atrophaeus (ATCC® 9372) 10 ^ 6 spores per strip.
Each Kit contains 3 Bionova BT60 Strip with Dual Spores.
Characteristics
The kit contains 1 Control Strip and 2 Test Strips.
25 mm x 70 mm non-absorbent material package, Violet code
5.0 x 25.0 mm filter paper strips.
10^5 or 10^6 Geobacillus stearothermophilus ATCC® 7953 spores and 10^6 Bacillus atrophaeus ATCC® 9372 spores per strip. Upon completion of the sterilization cycle, the spore strip should be tested in a suitable culture medium at 37 °C or at 60 °C.
We suggest the use of Bionova® MC1020-2 or MC1030-2 culture media to final reading in 48 hours at 37 °C for Ethylene Oxide and Dry Heat sterilization process control, or Bionova® MC20-2 or MC1020-2 culture media to final reading in 24 hours at 60 °C for Steam sterilization process control. For conventional culture media like TSB, incubation for 7 days is recommended.
D-ValueSTEAM: Not lower than 1.5 minutes. Conditions: 121 ºC.
Z-ValueSTEAM: Not lower than 6 ºC.
D-ValueEO: Not lower than 2.0 minutes. Conditions: 600 mg/l EO (100%), 60 % RH, 54 °C.
D-ValueDRY HEAT: Not lower than 2.0 minutes. Conditions: 160 °C.