Skip to content

Bowie-Dick Helix kit (metal)

SKU: KH2X15-3.5BD/M Category: Tag:
Additional information
Brand

Chemdye

Process

Steam

Packaging

1 metallic device + 1 cotton bag + 250 strips

Initial Color

Blue

Final Color

Black

Indicator Type

Type 2 Chemical Indicator

Regulations

EN 285, EN 867-5, ISO 11140-1

Description

KH2X15-3.5BD/M / Helix-PCD + Bowie-Dick simulation test kit for Steam sterilization processes: 3,5 min.134° C. Sterilizer monitoring. Metal capsule.Plastic tube of 1,5 m and 2 mm internal diameter. 1 cotton bag + 250 strips PCD-A-3.5BD Turns from Blue to Black. EN 867-5, ISO 11140-1 Type2. ISO 11140-4.

Designed to assess air removal and steam penetration inlarge and small vacuum-assisted steam sterilizers. Condition: Bowie & Dick cycle: 3.5 minutes, 134 ºC.

Characteristics

Helix-PCD

Metal capsule of great resistance and durability, to hold a Chemical Indicator designed for this system. The capsule is connected to a high resistance plastic tube of 1.5 m length (see catalogue for other lengths) and 2 mm internal diameter. Special silicone connector coated by external connector.

PCD-A-3.5BD

Type 2 Chemical Indicator.

Self-adhesive strip printed with indicator ink. Size: 76 x 6 mm.

10 strips per sheet. Size: 80 x 98.5 mm.

Initial color: blue.

Final color: dark grey/black.

Authorization: ANMAT (Argentinean National Administration of Drugs, Food and Medical Devices) PM-1614-4.

NOTE: If 4 bars change to the final color, the result is successful, the sterilizer works properly. If one or more bars have not turned to the final color, this shows a failure in sterilizing agent penetration, therefore an equipment malfunction.

Description

KH2X15-3.5BD/M / Helix-PCD + Bowie-Dick simulation test kit for Steam sterilization processes: 3,5 min.134° C. Sterilizer monitoring. Metal capsule.Plastic tube of 1,5 m and 2 mm internal diameter. 1 cotton bag + 250 strips PCD-A-3.5BD Turns from Blue to Black. EN 867-5, ISO 11140-1 Type2. ISO 11140-4.

Designed to assess air removal and steam penetration inlarge and small vacuum-assisted steam sterilizers. Condition: Bowie & Dick cycle: 3.5 minutes, 134 ºC.

Characteristics

Helix-PCD

Metal capsule of great resistance and durability, to hold a Chemical Indicator designed for this system. The capsule is connected to a high resistance plastic tube of 1.5 m length (see catalogue for other lengths) and 2 mm internal diameter. Special silicone connector coated by external connector.

PCD-A-3.5BD

Type 2 Chemical Indicator.

Self-adhesive strip printed with indicator ink. Size: 76 x 6 mm.

10 strips per sheet. Size: 80 x 98.5 mm.

Initial color: blue.

Final color: dark grey/black.

Authorization: ANMAT (Argentinean National Administration of Drugs, Food and Medical Devices) PM-1614-4.

NOTE: If 4 bars change to the final color, the result is successful, the sterilizer works properly. If one or more bars have not turned to the final color, this shows a failure in sterilizing agent penetration, therefore an equipment malfunction.