30min Biological Indicator for Steam (Gravity Displacement – IUSS) (FDA)

More info about this product

Additional information
Brand

Bionova

Process

Steam

Packaging

50

Microorganism

Geobacillus stearothermophilus (ATCC® 7953)

Population

10^6 Spores/carrier

Read-Out Time

30 min

Regulations

IRAM 37102-1, IRAM 37102-3, ISO 11138-1, ISO 11138-3

Description

Super Rapid Self-contained Biological Indicator for Steam sterilization processes. Geobacillus stearothermophilus (ATCC 7953) 10^6 spores per vial. Readout: 30 minutes. Please review the instructions for use for applicable cycles in your country

Terragene® Bionova® BT223 Fluorescence Super Rapid Readout Biological Indicators are single-use Self-Contained Biological Indicators (SCBI) that consist of a polypropylene tube, a spore carrier and a glass ampoule with a culture medium, enclosed with a colored cap.

Each tube contains a population of Geobacillus stearotermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Steam. Each BT223 has a Process Indicator on label that changes from pink to brown when exposed to Steam.

Instructions for Use

1. Identify the Terragene® Bionova® BT223 SCBI by writing the sterilizer number (in case of having more than one), load number and processing date on the label.

2. Pack the SCBI along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas which are considered most inaccessible for the sterilizing agent (e.g., the center of the load and areas near the door).

3. Sterilize as usual.

4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the SCBI from the package. CAUTION: Wear safety glasses and gloves when removing the Terragene® Bionova® BT223 SCBI from the sterilized package. WARNING: Do not crush or handle the SCBI excessively, since this might cause the glass ampoule to burst. Let the SCBI cool down until it reaches room temperature.

5. Check the Process Indicator on SCBI label. A color change to brown indicates that the SCBI has been exposed to Steam. IMPORTANT: This color change does not evidence the process effectiveness to achieve sterility. If the Process Indicator color has not changed, check the sterilization process.

6. Press the lid to seal the tube. Crush the ampoule contained in the SCBI with an individual ampoule crusher or with the ampoule crusher placed within the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature in a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the SCBI in the incubator.

IMPORTANT: Use a non-sterilized SCBI as a positive control at least once per day, when a sterilization cycle is run. The positive control ensures that correct incubation conditions were met; capability of medium to promote rapid growth; viability of spores has not been altered due to improper storage temperature, humidity or proximity to chemicals and proper functioning of Terragene® Bionova® Auto-Reader Incubators. Both, the positive control indicator and the processed indicator, should belong to the same batch.

7. Incubate the processed indicator and the positive control indicator in the appropriate Terragene® Bionova® Auto-Reader Incubators for a maximum of 30 minutes at (60 ± 2) °C for super rapid readout.

NOTE: Holding time between sterilization and incubation should not exceed a 7-day period. Fluorescence detection by the reader (excitation 340-380 nm / emission 455-465 nm) means a failure in the sterilization process. If no fluorescence is detected at 30-minute incubation, the result is negative. The positive control must give positive fluorescence readout. It is good practice to incubate a positive control for a visual color change.

Description

Super Rapid Self-contained Biological Indicator for Steam sterilization processes. Geobacillus stearothermophilus (ATCC 7953) 10^6 spores per vial. Readout: 30 minutes. Please review the instructions for use for applicable cycles in your country

Terragene® Bionova® BT223 Fluorescence Super Rapid Readout Biological Indicators are single-use Self-Contained Biological Indicators (SCBI) that consist of a polypropylene tube, a spore carrier and a glass ampoule with a culture medium, enclosed with a colored cap.

Each tube contains a population of Geobacillus stearotermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Steam. Each BT223 has a Process Indicator on label that changes from pink to brown when exposed to Steam.

Instructions for Use

1. Identify the Terragene® Bionova® BT223 SCBI by writing the sterilizer number (in case of having more than one), load number and processing date on the label.

2. Pack the SCBI along with materials to be sterilized in an appropriate package according to recommended sterilization practices. Place the package in those areas which are considered most inaccessible for the sterilizing agent (e.g., the center of the load and areas near the door).

3. Sterilize as usual.

4. After the sterilization process has finished, open the sterilizer door, wait five minutes and remove the SCBI from the package. CAUTION: Wear safety glasses and gloves when removing the Terragene® Bionova® BT223 SCBI from the sterilized package. WARNING: Do not crush or handle the SCBI excessively, since this might cause the glass ampoule to burst. Let the SCBI cool down until it reaches room temperature.

5. Check the Process Indicator on SCBI label. A color change to brown indicates that the SCBI has been exposed to Steam. IMPORTANT: This color change does not evidence the process effectiveness to achieve sterility. If the Process Indicator color has not changed, check the sterilization process.

6. Press the lid to seal the tube. Crush the ampoule contained in the SCBI with an individual ampoule crusher or with the ampoule crusher placed within the incubator´s incubation area. Then shake the tube down vigorously, with movements similar to those performed to lower the temperature in a mercury thermometer, until the medium reaches the base of the tube and soaks the spore carrier entirely. Finally, place the SCBI in the incubator.

IMPORTANT: Use a non-sterilized SCBI as a positive control at least once per day, when a sterilization cycle is run. The positive control ensures that correct incubation conditions were met; capability of medium to promote rapid growth; viability of spores has not been altered due to improper storage temperature, humidity or proximity to chemicals and proper functioning of Terragene® Bionova® Auto-Reader Incubators. Both, the positive control indicator and the processed indicator, should belong to the same batch.

7. Incubate the processed indicator and the positive control indicator in the appropriate Terragene® Bionova® Auto-Reader Incubators for a maximum of 30 minutes at (60 ± 2) °C for super rapid readout.

NOTE: Holding time between sterilization and incubation should not exceed a 7-day period. Fluorescence detection by the reader (excitation 340-380 nm / emission 455-465 nm) means a failure in the sterilization process. If no fluorescence is detected at 30-minute incubation, the result is negative. The positive control must give positive fluorescence readout. It is good practice to incubate a positive control for a visual color change.