Reusable Housing
The Helix Plus plastic housing can be used for up to 250 cycles.
화학 지표
신제품

Chemdye® Helix Plus is a Type 2 system with a reusable PCD for monitoring air removal and steam penetration in sterilization cycles involving narrow lumens.
Hollow, complex and minimally invasive devices are increasingly common in CSSDs. Inside narrow lumens, residual air, vacuum leaks or non-condensable gases may interfere with effective steam penetration — while quality and traceability requirements continue to rise. Helix Plus Kit helps CSSDs monitor these critical conditions through a renewed and connected Helix system.
HELIX KIT PLUS was developed for today's steam sterilization monitoring workflows, combining a renewed physical design with digital traceability tools.
Helix Plus is: A plastic housing incorporating a plastic capsule designed to hold a PCD-A chemical indicator specifically designed for this system. The capsule includes a screw cap with an indicator attached, and is connected at one end to a 1500 mm long, 2 mm internal diameter plastic tube. The plastic tube is arranged in a helical pattern and must not be released or removed from the housing to use the device.
250 PCD-A chemical indicator strips: Type 2 chemical indicator. Self-adhesive strip printed with indicator ink. Initial color: yellow. Final color: dark brown/black.
The Helix Plus plastic housing can be used for up to 250 cycles. The shelf life of the PCD-A chemical indicators is 5 years.
Choose the cycle time you need:
The Helix Plus plastic housing can be used for up to 250 cycles.
PCD-A strips are a Type 2 chemical indicator, self-adhesive and printed with indicator ink, turning from yellow to dark brown/black.
Designed under Quality Management System standards ISO 13485:2016/EN ISO 13485:2016, ISO 11140-1:2014, and ISO 11140-6:2022.
Available for steam sterilization cycles of 3.5, 5.3, or 7.0 minutes at 134 °C, with 5.3Y and 7.0Y variants also validated at 121 °C.
The shelf life of the PCD-A chemical indicators is 5 years.
Data Matrix identification. Compatible with Trazanto® Lens. Integrated with Bionova® Q App.

Sterilization loads are evolving. Monitoring must evolve with them